_
|
Capítulo 1: Página 12 – DC Characteristics
Cambiar: Continuity 24-Hour 90 Day 1 Year
0.002 + 0.010 0.008 + 0.020 0.010 + 0.020
A: Continuity 24-Hour 90 Day 1 Year
0.002 + 0.030 0.008 + 0.030 0.010 + 0.030
Capítulo 1: Páginas 12-13 - DC Characteristics and Measuring/Operating Characteristics
Agregar la nota [12] al pie en parámetro “Diode Test” de “Function” (página 12) y el siguiente texto a la
página 13:
“[12] Accuracy specifications are for the voltage measured at the input terminals only. 1mA test current is
typical. Variation in the current source will create some variation in the voltage drop across a diode junction."
Capítulo 1: Página 17 - Frequency/Period Measuring/Operating Characteristics
Cambiar: [4] Input > 100 mV. For 10 mV input, multiply % of reading error x10.
A: [4] Input > 100 mV. For 10 mV to 100 mV inputs multiply % of reading error x10.
Capítulo 1: Página 18 – General Information
Cambiar: Warranty: 3 years standard
A: Warranty: 1 year standard
Capítulo 1: Página 19 – Product Dimensions
1. Cambie las dimensiones en la parte superior del dibujo de la siguiente manera:
2. Ubique y cambie la dimensión del panel frontal de ‘42.58’ (mm) a ‘67.53’ (mm) en la parte superior
izquierda del diagrama inferior.
Capítulo 6: Página 112 - Types of Service Available
Cambiar: “If your multimeter fails within three years of original purchase…”
A: “If your multimeter fails within one year of original purchase…”
Cambiar: “If your unit fails after your three year warranty expires…"
A: “If your unit fails after your one year warranty expires…”