Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Installation Instructions for 33210 80001 Retrofit Kit Retrofit Instructions ...

Page 2: ...ghts customarily provided to the public to use modify reproduce release per form display or disclose commercial computer software or commercial com puter software documentation No additional government requirements beyond those set forth in the EULA shall apply except to the extent that those terms rights or licenses are explicitly required from all providers of commercial computer software pursu ...

Page 3: ...ng and Control Instrument product The affixed product label is as shown below Do not dispose in domestic household waste To return this unwanted instrument contact your nearest Keysight Service Center or visit http about keysight com en companyinfo environment takeback shtml for more information Sales and Technical Support To contact Keysight for sales and technical support refer to the support li...

Page 4: ...4 Keysight 33210A Retrofit Instructions THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 5: ...Retrofit Kit This guide shows how to install the Option 001 External Timebase Reference assembly into the Keysight 33210A This assembly can be installed on all Keysight 33210A instruments all serial numbers that were not shipped with Option 001 installed The Keysight 33210A Function Arbitrary Waveform Generator User s Guide contains instructions on using the features identified as option 001 only ...

Page 6: ...ternal Timebase Reference 6 Keysight 33210A Retrofit Instructions Installing the Timebase Assembly 1 Turn off the power Remove all cables from the instrument 2 Rotate the handle upright and pull off 3 Pull off the instrument bumpers ...

Page 7: ...ve the rear bezel 5 Slide off the instrument cover Loosen Screws Torx T 20 Slide Cover Off WARNING SHOCK HAZARD Make sure that the power cable has been removed from the instrument before proceeding Dangerous voltages may be encountered inside the instrument chassis at any time when the power cord is connected to power even with the power switch turned off ...

Page 8: ...m the chassis CAUTION Electrostatic discharge ESD can damage components on unprotected circuit boards Use static safe procedures when handling circuit boards Avoid high static areas such as carpeted floors If an anti static wrist strap is available its use is recommended Otherwise touch the metal instrument chassis to discharge static electricity before removing the timebase assembly from its anti...

Page 9: ...o the side 9 Remove the two plastic caps from the back panel The holes in the back panel for the 10 MHz In and 10 MHz Out connectors are capped in all instruments that do not have Option 001 installed Locate the plastic caps and from inside the chassis compress the retention tabs and push each cap out through the back panel Keyed connector Push out both plastic caps ...

Page 10: ...en them with the hex nuts and lock washers provided The installed assembly should be component side up as shown below c Tighten the hex nuts with a 9 16 inch open end or adjustable wrench to secure the assembly Do not overtighten just one quarter turn past hand tight 11 Reassemble the instrument by reversing steps 1 through 8 The Keysight 33210A is ready to use No software updates are required The...

Page 11: ...ion Installing the External Timebase Reference assembly has no effect on the instrument s calibration No recalibration is necessary unless it is time for a scheduled calibration Refer to the Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide for further information about calibration ...

Page 12: ...1 Keysight 33210A Option 001 External Timebase Reference 12 Keysight 33210A Retrofit Instructions THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 13: ...n is subject to change without notice Always refer to the Keysight website for the latest revision Keysight Technologies 2008 2017 Edition 2 October 1 2017 Printed in Malaysia 33210 90020 33210 90020 www keysight com ...