To disassemble the instrument
Use the following procedure to disassemble the 1143A Probe Offset Control
and Power Module.
W A R N I N G
Hazardous voltages exist on the power module. Failure to adhere closely to
the following procedures can cause electrical shock.
1
Remove the power cord.
2
Remove four flathead screws and remove the top cover.
W A R N I N G
Be sure to reconnect the safety ground when reassembling the instrument.
Failure to reconnect the safety ground can result in electrical shock.
3
Unplug the safety ground from the tab on the rear panel of the
instrument.
4
At the PC board, disconnect the cables that connect the front panel
probe power outputs.
5
Note the orientation of the knobs. Remove the four knobs.
6
On the bottom of the instrument, remove the 5 mm screw that fastens
the transformer support.
7
Remove the following pan-head screws.
•
Three directly on the PC board.
•
Two on the ac input connector.
•
Two on the heatsink.
8
Remove the PC board. Slide it slightly forward so parts will clear the
rear panel, then lift the rear of the board out while sliding it
backwards.
9
Remove the two heatsink spacers from the standoffs that were
directly under the hearsing.
10
Reverse the procedure to reassemble the power module.
Service
Troubleshooting and Repair
34
Summary of Contents for 1143A
Page 8: ...8 ...
Page 16: ...16 ...
Page 18: ...18 ...
Page 43: ...Figure 6 Power and Control Assembly Component Locator Service Troubleshooting and Repair 43 ...
Page 46: ...46 ...
Page 48: ...Figure 8 Schematic 1 Line Input and Positive Supply 1 Service Troubleshooting and Repair 48 ...
Page 49: ...Figure 9 Schematic 2 Negative Supply 2 Service Troubleshooting and Repair 49 ...
Page 52: ...52 ...