Oscilloscope Utility Settings
6
Keysight 1000B Series Oscilloscopes User’s Guide
137
Running Self-Calibration
The automatic calibration routine adjusts the internal circuitry of the oscilloscope
for the best measurement accuracy.
The automatic calibration should be run when the ambient temperature changes
by 5
°
C or more.
To run the oscilloscope’s self-calibration:
1
Press
[Utility]
.
2
In the Utilities menu, press
Self-Cal
.
3
Follow the instructions on the Calibration screen.
NOTE
Before performing the automatic calibration, let the oscilloscope warm-up at least 30
minutes.
Figure 58
Calibration Screen