Keysight U4205A Direct Connect 34-Channel Mictor Probe
25
Circuit Board Design
4
16850 Series and U4154A Data and Clock Inputs per Operating Mode
The following table shows the number of data and clock inputs for each
connector on your target system for the various operating modes of your
U4154A and 16850 Series logic analyzer.
Table 4
Data clock inputs per operating mode
U4154A and 16850 Series
Operating Mode
U4205A
Synchronous (state)
analysis maximum
800 Mb/s
32 data
plus 2
clock inputs
(see Note below)
Eye scan mode 800 Mb/s
32 data
plus 2
clock inputs
(see Note below)
Timing mode
32 data
plus 2
clock inputs
NOTE
In synchronous (state) analysis mode, and in the eye scan mode, there is one
clock input which must be routed to the clock input on pod 1. The clock inputs
on other pods can be assigned to labels and acquired as data inputs.