10
Keysight U3020AS22 User’s and Service Guide
U3020AS22
Controlling the Test Set
1-
Controlling the Test Set
Controlling the Test Set and Making Measurements
Keysight U3020A is a “slave” instruments. A controller must be used to control
the Test Set. There are two methods that can be used to control the Test Set.
— Using LAN connection
— Using GPIB connection
Once the connection between the Controller and the Test Set has been
established (LAN or GPIB), the Test Set can be controlled using SCPI
commands.
LAN Connection using Your Company’s Intranet
Connect the Test Set and your controlling computer to an active instrument
LAN line. Open the Internet browser application on the computer. Enter this
default web address on your browser application: http://U3020AS22-xxxx
The Welcome screen for the instrument should be viewable if the LAN
connection is working. Select the following screen control buttons to access
the SCPI command control window.
1.
Browser Web Connect (control button on the left side of the screen)
2.
Allow Full control
3.
Commands
The last four digits in the web address (xxxx) are the last four digits of the
instruments serial number.
Once the connection between the controller and the Test Set has been
established (LAN or GPIB), the test set can be controlled using SCPI
commands. Refer to
.