Keysight U1701B Quick Start Guide
9
Keysight U1701B Quick Start Guide
9
ENGL
ISH
To cycle through 1%, 5%, 10% and 20% tolerance
–
will be indicated.
– Beeper will beep once if the test value is within the
selected tolerance. If the test value is out of the
tolerance, the beeper will beep three times.
– This mode cannot be enabled under the following
conditions:
– After setting the recording mode
– After setting the compare mode
– Display showing either
OL
or below 10 counts
Press
momentarily
To enable compare mode
– Measuring range will be locked
–
will be displayed and the secondary display
will indicate
C
# #
, meaning which set has been
used for compare mode. The two right digits
indicate current compare set. The
# #
range from
01 to 25.
– The primary display shows the present
measurement. In this state, it is ready for testing.
– If the reading is beyond the high limit,
will be
indicated.
will be indicated if the reading is
out of the low limit. The beeper will beep three
times and the secondary display will indicate
nGo
.
– If the reading is within the high and low limits, the
beeper will beep once, and the secondary display
will indicate
Go
. After three seconds or when the
reading is lower than 10 counts, the instrument
will return to its ready state.
– The secondary display will indicate
C01
to
C25
according to the comparison record that has been
selected.
Press
To save comparison set for next entry
Press
and
hold for more than 1 s
Actions
Steps