DI/DO (M8) Connector
In Safety Disconnect mode the DI/DO inputs and outputs are intended to control an auxiliary AC
contactor.
Rear panel pinout view
Typical connections
Pin
Description
1
DOutput, +24V drive signal for AC contactor (drive current ≤500 mA)
2
DOutput, 24V drive signal return
3
DInput, AC contactor OK when shorted to pin 4
4
DInput, AC contactor OK when shorted to pin 3
Mating Connector Part Numbers
Connector
Mates with
Phoenix Contact part number
M12 female
Remote Interface (M12)
1404420 (Wire gauge 24-28 AWG)
M8 female
Fixture Cover (M8)
1681185 (Wire gauge 22-28 AWG)
M8 male
DI/DO (M8)
1681169 (Wire gauge 22-28 AWG)
Typical Customer-supplied Components
Component
Description
Part number
ESTOP switch
2-pole NC, latching, compatible with 24V
Eaton C22-PVT45P-K02
RESET lamp/switch
Blue LED 24V with momentary contact NO switch
Eaton C22S-DL-B-K10-24
POWER ON lamp
Green LED 24V
Eaton C22-L-G-24
RELAYS CLOSED lamp
Yellow LED 24V
Eaton C22-L-Y-24
Safety cover switch
Mates directly with a magnetic relay
Pilz magnetic relay 504222
AC contactor
3-phase AC contactor with NC auxiliary contacts
Omron J7KNA-09-01
Appendix A Keysight SD1000A Safety Disconnect System