Keysight Double-Pulse Test Rack and Safety Enclosure Operation Guide
Operating the Rack and Safety Enclosure
Using the Rack and Safety Enclosure
Using the Rack and Safety Enclosure
Main On/Off Switch
The Rack and Safety Enclosure is supplied by one main AC voltage input. Turn it
on or off with the main power switch at the front of the rack.
1
Turn the power switch to the
OFF
position.
2
Make sure
Emergency Stop
is pulled out.
3
Turn the power switch to the
ON
position. The internal fans start, the red and
amber lights (in the light tower) turn on.
4
Press the
Reset
button. You should hear a few clicks as the safety relays
energize. The amber light turns off followed by the red light. The green light
turns on.The system is now operational.
To power off the complete system, turn the power switch to
OFF
.
Power Switch On
Power Switch Off
In general, you should turn off the PD1500A solution when it will
not be used for long periods of time. This is to prevent a problem
with the Heinzinger HV power supply. If the solution has been left
on and cannot communicate with the HV power supply, do the
following:
1. Shut down the PD1500A DPT rack.
2. Disconnect the PD1500A from the ac power socket.
3. Wait a few minutes.
4. Reconnect the PD1500A rack to the ac power socket.
5. Turn on the PD1500A.