Keysight Double-Pulse Test Rack and Safety Enclosure Operation Guide
19
Rack Ventilation
Introduction
Rack Ventilation
The Rack and Safety Enclosure has four air intake fans (two on each side) and
two large exhaust fans on the rear of the Safety Enclosure hood. Additionally,
there are two vents (no fans) on the rear of the rack for warm air exhaust.
For proper air flow through the Rack and Safety Enclosure, do not
block air vents or rack cooling fans. Allow a minimum of 50 cm on
each side and on the rear of the Rack and Safety Enclosure for
ventilation.
Figure 2
Air Flow through the DPT Rack and Safety Enclosure
The Lower Air Intake fans on the chassis sides are temperature
controlled. They turn on only when the temperature in the lower
portion of the rack exceeds approximately 30
°
C.
Air exhaust vents
from Safety
Enclosure
(two fans inside
safety hood)
Air exhaust vents
from Rack and
Enclosure (no fans
inside)
Upper Air
intake fans.
One on each
side.
Lower Air
intake fans.
One on each
side of Rack
and Safety
Enclosure. See
Note below.