22 Keysight NFA X-Series Getting Started and Troubleshooting Guide
Quick Start
Using the U7227-Series USB Preamplifiers in Noise Figure Measurements
Initial Connection to the Signal Analyzer
The U7227-Series USB Preamplifier is powered via a USB connection from the
signal analyzer. To connect it to the analyzer, follow the steps below.
1.
Connect the Output port of the USB preamplifier to the RF Input port of
the signal analyzer.
2.
Connect the USB port (Type-B) of the USB preamplifier to the USB port
(Type-A) of the signal analyzer either in the front panel or in the rear panel
using a USB cable. You can “hot plug” the USB preamplifier into the
analyzer’s USB port at any time.
shows the U7227-Series USB Preamplifier connection in the two
stages of the noise figure measurement. While the USB preamplifier is plugged
into one of the analyzer’s USB ports, the analyzer will consider it to be in the
RF input signal path and will apply the gain settings and other calibration data
of the USB preamplifier to the measurements.
Figure 1-2
U7227-Series USB Preamplifier and Analyzer Connection in the Two Stages of
Noise Figure Measurement
Before connecting a signal to the input port of the U7227-Series USB
Preamplifier, make sure the USB preamplifier can safely accept the signal
level provided. The signal level limits are marked next to the Input
connectors of the USB preamplifier.
The USB connections of the U7227-Series USB Preamplifier and the
keyboard to the signal analyzer must be separated from each other; for
example, the USB preamplifier connects through the front panel while the
keyboard connects through the back panel.
Summary of Contents for NFA X-Series
Page 4: ...4 ...