175
Instrument Messages
Command Errors
Command Errors
The messages listed below are command errors, which are saved
into local registers.
-100
Command error
An unrecognized command or data type was encountered. This
message is used when the device cannot detect more specifics
described for errors -101 to -109.
-101
Invalid character
A syntactic element contains an invalid character.
-102
Syntax error
An unrecognized command or data type was encountered.
-103
Invalid separator
The parser was expecting a separator and encountered an illegal
character.
-104
Data type error
The parser recognized a data element that is not allowed.
-108
Parameter not allowed
More parameters were received than expected for the header. For
example, the *ESE common command only accepts one
parameter, so receiving *ESE 0,1 is not allowed.
-109
Missing parameter
Fewer parameters were received than required for the header. For
example, the *ESE common command requires one parameter, so
receiving *ESE is not allowed.
-110
Command header error
An error was detected in the header. This message is used when
the device cannot detect more specifics described for errors -111
to -119.
Summary of Contents for N9321C
Page 36: ...Getting Started Helpful Tips 28 ...
Page 110: ...Functions and Measurements Files Operation 102 ...
Page 180: ...Key Reference Trace Det 172 ...
Page 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Page 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Page 216: ...Menu Maps Trace Det Menu 208 ...
Page 219: ......