background image

Notice: This document contains references to Agilent. 
Please note that Agilent’s Test and Measurement 
business has become Keysight Technologies. For more 
information, go to www.keysight.com

.

 

Keysight Technologies

N9040B UXA Signal Analyzer

Option B2X, 255 MHz Analysis Bandwidth

Installation Note

Summary of Contents for N9040B

Page 1: ...Agilent Please note that Agilent s Test and Measurement business has become Keysight Technologies For more information go to www keysight com Keysight Technologies N9040B UXA Signal Analyzer Option B2X 255 MHz Analysis Bandwidth Installation Note ...

Page 2: ...implied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Manual Part Number N9040 90017 Supersedes None Edition Edition 1 April 2016 Printed in USA Malaysia Published by Keysight Technologies Inc 140...

Page 3: ...lyzer The option is licensed for one instrument model number serial number combination The license file that is downloaded from the web will only install on the designated instrument Products Affected UXA N9040B Serial Numbers All To Be Performed By X Keysight Service Center Personnel Qualified by Keysight Customer Estimated Installation Time Estimated Adjustment Time Estimated Verification Time 2...

Page 4: ... Analyzer Service Guide N9040 90006 Available online Quantity Description Keysight Part Number 1 Installation Note This note 1 Option Upgrade Entitlement Certificate 5964 5178 1 Wideband Analog IF assembly N9020 60043 1 Wideband Digital IF assembly N9020 60311 1 Flex Circuit WB DIF to Flex Circuit Interconnect Bracket N9040 60045 1 Flex Circuit WB AIF to Flex Circuit Interconnect Bracket N9040 600...

Page 5: ...the instrument and allow the instrument to boot up run the alignments and display the measurement screen The instrument will probably display a spectrum analyzer screen and you will see the instrument sweeping 2 There should be no alignment failures If there are failures investigate and fix the problem before continuing ...

Page 6: ...band Analog IF to be installed 4 Locate the Wideband Analog IF Assembly and the N9040 60042 Flex Circuit Cable in the kit Install the Flex Circuit cable into the connector on the Wideband Analog IF Assembly This flex cable will be connected to the other flex cable in the kit later in the installation process 5 Install the Wideband Analog IF into card cage connectors J8 and J10 See Figure 1 6 Locat...

Page 7: ...nce assembly at J718 10 Locate W41 cable with color code 17 at one end and color code 901 at the other end Attach the color code 17 end to the Wideband Analog IF at J17 and the other end to A15 Front End Controller at J901 Remove cable that is on A15J901 it will be discarded after the other end of the cable at A3J15 is removed later 11 Locate W40 cable with color code 18 at one end and color code ...

Page 8: ...ion Note N9040 90017 Option B2X 255 MHz Analysis Bandwidth 13 See Figure 3 that shows where the color code 15 end of W40 connects to the A3J15 Digital IF assembly Figure 3 DIF Cables Showing W40 Connection to J15 ...

Page 9: ...t cable in the kit and connect one end to the Wideband Digital IF assembly and the other end to the Flex Cable previously installed on the Wideband Analog IF assembly 16 Locate the Bracket Flex Circuit Interconnect in the kit This bracket secures the flex circuit connection you made in the previous step See Figure 4 attach the flex cable connection to the interconnect bracket using the two flat he...

Page 10: ...flat head screw 0515 1946 is required to secure the new Wideband Digital IF assembly to the rear panel This flat head screw is in the bag of screws in the kit Torque to 9 inch pounds 19 Re install the front top brace and use the 0515 1946 flat head screws in the kit in the countersunk locations Torque to 9 inch pounds 20 Reinstall the rear top brace and outer case side strap handles and feet 21 Po...

Page 11: ...virus scan on this device before use 5 Save the License File to the root directory of the USB storage device 6 Connect the USB storage device to the signal analyzer USB port Windows will detect the new hardware and may display the configuration menu shown in Figure 5 This menu may be configured according to your preferences Figure 5 USB Storage Device Configuration Menu 7 The signal analyzer will ...

Page 12: ...am Files Agilent licensing Verify the License Installation and Hardware 1 Cycle power on the signal analyzer and wait until the analyzer boots to the measurement application screen 2 Press System Show System to display a list of installed options 3 Verify that the installed options list contains the newly installed N9040B B2X You will see Option DP4 appear also and this is the extra capture memory...

Page 13: ...1 Press MODE MEAS and select I Q Analyzer Basic Assure Complex Spectrum is highlighted Tap OK Tap Frequency to view the pull down menu and select Meas Setup Tap IF Path then IF Path again and verify 255 MHz appears Select this 255 MHz path 2 Turn on the 50 MHz calibrator signal Input Output RF Calibrator select 50 MHz and set the span to 255 MHz press Frequency select Span 255 MHz Tune the analyze...

Page 14: ...ffice To find your local Keysight office access the following URL or if in the United States call the following telephone number http www keysight com find assist 1 800 829 4444 8 am 8 pm ET Monday Friday Adjustment Name Perform all IF Flatness adjustments Remember to perform YTF Alignment Press System Alignments Advanced Characterize Preselector Perform Characterize Noise Floor Press System Align...

Page 15: ...This information is subject to change without notice Keysight Technologies 2016 Edition 1 April 2016 N9040 90017 www keysight com ...