12
SYSTem Subsystem
524
Keysight N8262A Programming Guide
NADC
The NADC set-up provides:
– Average power measurement of both active timeslots in NADC or IS-136 “full
rate” transmission. This assumes that there are two timeslots to be measured
in each frame as for example with timeslots 0 in the following diagram:
Figure 12-2
A trace display of the active timeslots
– A trace display of the active timeslots.
The measurement is started by detecting the RF burst—for example the burst
emitted by a mobile—using the internal RF level trigger. The internal level trigger
is set to –20 dBm. Time-gating is used to measure the average power in two
active timeslots which are separated by two inactive timeslots
Table 12-40
TD-SCDMA: Power meter presets: Window/Measurement settings
Function
Secondary Channel Sensor
No Sensor
Non P-Series Sensor
P-Series Sensor
Display setup
Lower window
Dual numeric
Dual numeric
Dual numeric
Lower window/lower measurement (LL)
Feed
Gate 1 primary channel
[a]
Secondary channel
[a]
Gate1 secondary channel
[a]
(Channel B)
Measurement
Pk-to-Avg
Avg
Avg
[a] For further information refer to
“Primary and Secondary Channels”
0
1
2
0
2
1
IS-136 full rate frame
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...