PSTatistic Subsystem
9
Keysight N8262A Programming Guide
329
PSTatistic[1]|2:CCDF:DATa?
This command is used to return 501 probability values in % at different power
level within certain range, starts from 0 dB till the predefined maximum power
level.
Syntax
Example
NOTE
The maximum power level can be set by using this command:
PSTatistic[1]|2:CCDF:DATa:MAX <numeric_value>
By default, the maximum value is 50 dB.
The power interval between each reading (probability value) is determined by
the defined maximum power level divided by 500.
NOTE
This command is only applicable when P-Series sensors are present and the
following conditions are met:
– Free run acquisition mode is selected
–
NORMal
or
DOUBle
measurement speed setting is chosen
PST
2
:CCDF
:DAT
1
?
PST1:CCDF:DAT?
This command returns 501 probability values
in % at different power levels within certain
range (from 0 dB to maximum power level
defined).