7
MEMory Subsystem
306
Keysight N8262A Programming Guide
For sensosr calibration tables (8480 Series sensors only), the number of frequency
points must be one less than the number of calibration factor points. This is
verified when the sensor calibration table is selected using
SENSe:CORRection:CSET:SELect <string>
.
Ensure that the frequency points you use cover the frequency range of the signals
that you want to measure. If you measure a signal with a frequency outside the
frequency range defined in the table, then the power meter uses the highest or
lowest point in the table to calculate the calibration factor/offset.
Depending on available memory, the power meter is capable of storing 20 sensor
calibration tables and 10 frequency dependent offset tables, each containing 80
points.
Syntax
Parameters
MEM
:TABL
:FREQ
numeric_value
Space
?
,
Item
Description/Default
Range of Values
numeric_value
A numeric value for the frequency. The default units
are Hz.
1 kHz to 1000.0 GHz
[a],[b]
[a] The following measurement units can be used:
Hz
kHz (10
3
)
MHz (10
6
)
GHz (10
9
)
[b] All frequencies are truncated to a multiple of 1 kHz.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...