MEMory Subsystem
7
Keysight N8262A Programming Guide
305
MEMory:TABLe:FREQuency <numeric_value>{,<numeric_value>}
This command is used to enter frequency data into the current selected table. Any
previous frequency list is cleared before the new frequency list is stored. The
frequencies must be entered in ascending order. Entries in the frequency lists
correspond as shown in
with entries in the calibration/offset factor lists.
NOTE
For sensor calibration tables only, the first calibration factor entered using the
MEMory:TABLe:GAIN
command is used as the reference calibration factor.
Table 7-3
Frequency and Calibration/Offset factor list
Table
Power Sensor
Table Name
0
None
DEFAULT
[a]
1
8481A
8481A
2
8482A, 8482B, 8482H
8482A
3
8483A
8483A
4
8481D
8481D
5
8485A
8485A
6
R8486A
R8486A
7
Q8486A
Q8486A
8
R8486D
R8486D
9
8487A
8487A
[a] Default is a sensor calibration table in which the reference calibration factor and calibration factors are 100%. This sensor calibration table
can be used during the performance testing of the power meter.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...