1
Specifications
34
Keysight N8262A Service Guide
1 mW Power Reference
Front panel inputs/outputs
NOTE
The 1mW power reference is provided for calibration of E-Series (E4410 and
E9300) and 8480 Series Sensors. The P-Series sensors are automatically
calibrated do not need this reference calibration.
Power output
1.00 mW (0.0 dBm) - Factory set
±
0.4% traceable to the National
Physical Laboratory (NPL) UK
Accuracy (over 2 years)
±1.2% (0 to 55
o
C)
±
0.4% (25 to 10
o
C)
Frequency
50 MHz nominal
SWR
1.08 (0 to 55
o
C)
1.05 typical
Connector type
Type N (f), 50
Recorder outputs
[a]
[a] Two recorder outputs are available on the N8262A P-Series modular power meter.
Analog 0 to 1 V, 1 k
output impedance, SMB connector.
There are two recorder outputs with SMB connector.
Trigger input
Input has TTL compatible logic levels and uses a SMB connector.
Trigger output
Output provides TTL compatible logic levels and uses a SMB
connector.