Repair Guide
6
Keysight N8262A Service Guide
111
Use a 1.5 Hex Allen tool to disconnect the
trigger board from the mother board assembly.
Remove all the locks and flat washers.
Note:
You will be able to remove the trigger interface
board assembly at this step.
– Step 1: Lift and remove the top coaxial cable
labeled 2 (channel B) from J2.
– Step 2: Lift and remove the bottom coaxial
cable labeled 3 (channel B) from J3.
– Step 3: Lift and remove the top coaxial cable
labeled 4 (channel A) from J4.
– Step 4: Lift and remove the bottom coaxial
cable labeled 5 (channel A) from J5.
Note:
These coaxial cables originally routed under the
measurement board assembly.
Instructions
Visual
4
1
2
3