1. N678xA SMU connector
2. Guard shield (can be the
shield of a coaxial cable)
Multiple Load Wiring
If you are using local sensing and are connecting multiple loads to one output, connect each load to
the output terminals using separate load wires as shown.
1. Twist leads
2. 4-wire disabled
(indicator is off)
This minimizes mutual coupling effects and takes full advantage of the power module's low output
impedance. Keep each wire-pair as short as possible and twist or bundle the wires to reduce lead
inductance and noise pickup. The goal is to always minimize the loop area or physical space between
the + and - load wires from the power analyzer to the load. Keep the load leads under 14.7 meters (50
feet) per lead because of inductance effects
Note that Keysight Models N678xA SMU have additional wiring restrictions as previously discussed
under
.
If load considerations require the use of distribution terminals that are located away from the
instrument, connect the output terminals to the remote distribution terminals by a pair of twisted or
bundled wires. Connect each load to the distribution terminals separately. Remote voltage sensing is
recommended under these circumstances. Sense either at the remote distribution terminals or, if one
load is more sensitive than the others, directly at the critical load.
2 Installation
60
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......