9.
Select or clear the
"
Route PNA RF to rear panel
'
RF OUT
'
” check box as
indicated in
10.
Set the ALC mode according to the following information:
For most applications the ALC mode should be on “Enable Test Set RF
ALC” box
checked. For this condition “Max Power Limit at Multiplier’s RF
IN” is automatically
set to 11 dBm. The hardware ALC’s in the test set
provide a nominal
11 dBm power level at the RF inputs to the extender
modules. This applies
to both the RF and LO inputs. A nominal cable loss
between the test set
and extender modules is assumed.
If the ALC mode “Enable Test Set RF ALC” is off, the max power may be set
by the user in the “Max Power Limit at Multiplier’s RF IN” dialog box. If the
user adjusts the analyzer’s RF power, the RF power into the
millimeter-wave modules will change, but will not exceed the specified
max power in the dialog box. It is recommended that ALC “Enable Test Set
RF ALC” be turned off in Pulse mode. Without ALC the power control is
less accurate.
11.
Select OK. The analyzer will preset for millimeter operation after you
select OK in the next dialog window.
Figure 1-28
Millimeter System Selections
Select the “Standard PNA” configuration to return to normal PNA operation
without the heads.
Summary of Contents for N5261A
Page 1: ...User s and Service Guide Keysight Technologies N5261A and N5262A Millimeter Head Controller...
Page 6: ......
Page 7: ...Keysight N5261 62A User s Service Guide 1 Installation System Operation...
Page 54: ...48 Keysight N5261 62A User s Service Guide...
Page 55: ...Keysight N5261 62A User s Service Guide 49 Service Information...