Installation Note N5225-90117
39
Description of the Upgrade
Installation Procedure for the Upgrade
1.
To start the Keysight License Manager, press Start > Keysight License
Manager > Keysight License Manager . A Keysight License Manager dialog
box will appear.
2.
Right click the on the desired option and click
Delete
.
3.
In the Select Desired Option list, click
201
.
4.
In the Keysight License Manager dialog box that appears, press or click
Yes
to confirm delete.
5.
A message displays stating that the option removal was successful.
Step 33. Enable Options 401
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The network analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
— Refer to the license message you received from Keysight: Verify that the
analyzer’s model and serial numbers match those on the license message
you received from Keysight.
Option Enable Procedure
1.
Locate the email(s) from Keysight which contain license file attachments.
These emails are a result of
“Step 1. Obtain a Keyword and Verify the
.
2.
Copy the license file(s) from the email(s) to the
root directory
More than one license file may be copied to the USB flash drive.
3.
Connect the USB flash drive to the PNA, within 5 seconds, the PNA should
display a small "New licenses installed" message.
For this step, you will need a USB flash drive.
A single file may contain more than one feature.