Installation Note N5225-90105

                                                                                                                        

19

Step 13. Position the Cables and Wires to Prevent Pinching

On the top side of the PNA, carefully position the grey flex cables so they can’t be pinched between the covers 

and the rails.
On the bottom side of the PNA, carefully fold or push down the ribbon cables and wires so they can’t be 

pinched between the hardware and the outer cover. Ribbon cables and wires must never be positioned on top 

of hardware.

Step 14. Reinstall the Outer Cover

For instructions, click the Chapter 7 bookmark “Removing the Covers” in the PDF Service Guide

1

.

Step 15. Enable Option 401

Procedure Requirements

• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A keyboard must be connected to the network analyzer.

Option Enable Procedure

1. To start the option enable utility, press UTILITY 

, then 

, then 

. An 

option enable dialog box will appear.

2. Click the arrow in the 

Select Desired Option 

box. A list of available options will appear.

3. In the 

Select Desired Option

 list, click 

401 - Configurable Test Set

.

4. Using the keyboard, enter the license key in the box provided. The license key is printed on the the license 

message you received from Keysight. Enter this key 

exactly

 as it is printed on the message.

5. Click 

Enable

.

6. Click 

Yes

 in answer to the displayed question in the 

Restart Analyzer?

 box.

7. When the installation is complete, click 

Exit

.

Option Verification Procedure

Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer’s 

Help

 menu, click 

About Network Analyzer

.

2. Verify that “401” is listed after “Options:” in the display. Click 

OK

.

NOTE

If Option 401 has not been enabled, perform the 

”Option Enable Procedure”

 again. If the 

option is still not enabled, contact Keysight Technologies. Refer to 

“Getting Assistance 

from Keysight” on page 3

.

System

Service

Option Enable

Summary of Contents for N5224AU-401

Page 1: ...sight Add Configurable Test Set Upgrade Kit Installation Note To Upgrade PNA N5224A or N5225A Option 400 to Option 401 Upgrade Kit Order Numbers N5224AU 401 or N5225AU 401 Keysight Kit Number N5225 60...

Page 2: ......

Page 3: ...al item as defined in FAR 2 101 a or as Restricted computer software as defined in FAR 52 227 19 June 1987 or any equivalent agency regulation or contract clause Use duplication or disclosure of Softw...

Page 4: ...m Keysight By internet phone or fax get assistance with all your test and measurement needs Contacting Keysight Assistance with test and measurements needs and information on finding a local Keysight...

Page 5: ...rade adjustments and full instrument calibration To view the equipment list click the Chapter 3 bookmark Tests and Adjustments in the PDF Service Guide1 License Key Redemption NOTE The enclosed Option...

Page 6: ...rvice Guide is displayed scroll through the Contents section bookmarks to locate the information needed Protecting Your Workspace from Electrostatic Discharge For information click on the Chapter 1 bo...

Page 7: ...et to 9 in lbs 1 02 N m 1 N A T 20 TORX driver set to 21 in lbs 2 38 N m 1 N A 5 16 in 8 mm nutsetter or open end torque wrench set to 10 in lbs 1 13 N m 1 N A Products affected N5224A and N5225A Opti...

Page 8: ...45 20099 W21 RF cable A29 port 1 receiver coupler to A37 reference mixer switch 1 N5245 20110 W22 RF cable A33 port 1 coupler to front panel Port 1 CPLR ARM 1 N5245 20014 W23 RF cable A30 port 3 recei...

Page 9: ...B 1 N5245 20042 W41 RF cable A37 reference mixer switch to front panel REF 1 SOURCE OUT 1 N5245 20006 W42 RF cable REF 1 RCVR R1 IN to A37 reference mixer switch 1 N5245 20007 W43 RF cable A37 refere...

Page 10: ...yword and Verify the Information Step 2 Remove the Outer Cover Step 3 Remove the Front Panel Assembly Step 4 Remove Some Existing Cables Step 5 Remove a 3 dB Pad Step 6 Assemble the A37 Reference Mixe...

Page 11: ...t match those on your license key you will not be able to install the option If this is the case contact Keysight for assistance before beginning the installation of this upgrade Refer to Contacting K...

Page 12: ...8 A32 port 2 receiver coupler to A27 mixer brick R2 W111 N5224 20013 A29 port 1 receiver coupler to A33 port 1 coupler W112 N5224 20015 A30 port 3 receiver coupler to A34 port 3 coupler W113 N5224 200...

Page 13: ...7 7759 using two screws 0515 0372 Use a T 10 TORX driver to tighten the screws New parts are listed in Table 1 on page 7 Step 7 Install the A37 Reference Mixer Switch Assembly Refer to Figure 1 for th...

Page 14: ...To see an image showing the location of these cables click the Chapter 6 bookmarks Bottom RF Cables 4 Port Configuration Option 401 in the PDF Service Guide1 New parts are listed in Table 1 on page 7...

Page 15: ...2 port 2 receiver coupler to front panel port 2 SOURCE OUT W32 N5245 20097 Port 2 CPLR THRU to A36 port 2 coupler W28 N5245 20096 Port 4 CPLR THRU to A35 port 4 coupler W33 N5245 20108 A32 port 2 rece...

Page 16: ...2 RCVR R2 IN to A27 mixer brick R2 W39 N5245 20038 Port 4 RCVR D IN to A28 mixer brick D W45 N5245 20021 REF 4 RCVR R4 IN to 3 dB pad on A28 mixer brick R4 W30 N5245 20018 A35 port 4 coupler to front...

Page 17: ...coupler to front panel Port 1 SOURCE OUT W42 N5245 20007 REF 1 RCVR R1 IN to A37 reference mixer switch W41 N5245 20006 A37 reference mixer switch to front panel REF 1 SOURCE OUT W43 N5245 20009 A37...

Page 18: ...A New parts are listed in Table 1 on page 7 1 From the back side of the front panel use a blunt object in the cutouts in the lower front dress panel to push on the old overlay item and separate it fro...

Page 19: ...ted in Table 1 on page 7 1 Remove the protective backing from the new front panel overlay N5227 80005 item 2 Starting from either side loosely place the overlay in the recess on the lower front panel...

Page 20: ...utility press UTILITY then then An option enable dialog box will appear 2 Click the arrow in the Select Desired Option box A list of available options will appear 3 In the Select Desired Option list...

Page 21: ...structions click the Chapter 3 bookmark Tests and Adjustments in the PDF Service Guide1 If you experience difficulty with the basic functioning of the analyzer contact Keysight Refer to Contacting Key...

Page 22: ...This information is subject to change without notice Keysight Technologies 2011 2014 August 2014 N5225 90105 N5225 90105 www keysight com...

Reviews: