89600 VSA Software Configuration
Interleaved mode can be enabled by opening the Input > Extensions... window from
the menu bar. Then select the Enable Interleaving option, changes are applied
immediately.
External Clock
The sampling rate can be tuned thanks to the external clock. This clock must be
continuously present if the mode is selected otherwise an error will occur.
The range of clock input frequency is model dependent, as shown in the table below:
Models
Option
Frequency Range
M9710A
-SR4
5 GHz
To use this option, first connect an external clock signal in the accepted range to the
CLK IN connector on the digitizer front panel. Please, refer to the your digitizer
datasheet for external clock specification.
Then open the Input > Extensions... window from the menu bar. Select the Enable
external clock option, and double click the External clock frequency value to bring up
the editing window. You may then set the value of the external clock frequency being
applied. As shown below:
12
M9710A Hardware Extension of 89600 VSA Software