M9403A and M9404A Operational Check Procedure
Equipment Test Configuration Option II (Signal Source & Power Meter)
1. Adjust the Signal Source Stimulus Power for the M9403A-H02 to -30 dBm, or -10 dBm for
the M9403A-H01.
2. Set the Signal Source to a CW frequency of 50 MHz.
3. Perform a Power Meter calibration.
4. Connect the Stimulus Source to M9403A Access Port.
5. Connect the Power Meter Sensor to M9404A Access Port.
NOTE
Depending on your instrument’s power handling capability it may be necessary to
use a 20 dB attenuator to protect the instrument.
6. Connect optical ports between M9403A and M9404A using an optical cable.
7. Measure the Power and compare Link Loss (Pmeasured - Pstimulus).
8. Set the Signal Source to a CW Frequency of 10 GHz.
9. Repeat
thru
10.Set the Signal Source to a CW Frequency of 26.5 GHz.
11.Repeat
thru
12.Set the Signal Source to a CW Frequency of 50 GHz.
13.Compare the Link Loss to
.
NOTE
On the M9403A or M9404A H02 you should expect to see a 30 dB improvement
over the Link Loss. Return Loss may be different as well.