:MEASure Commands
21
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
477
:MEASure:DUAL:VPP
Overview
This measurement is available with the N2820A high sensitivity current probe
when both the Primary and Secondary probe cables are used. This measurement
joins the Zoom In waveform data below the probe's clamp level with Zoom Out
waveform data above the probe's clamp level to create the waveform on which the
measurement is made.
Command Syntax
:MEASure:DUAL:VPP [<source1>][,<source2>]
<source1>,<source2> ::= CHANnel<n> with N2820A probe connected
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:DUAL:VPP command installs a screen measurement and starts a
vertical peak-to-peak measurement.
If the optional source parameter(s) are specified, the currently specified source(s)
are modified.
Query Syntax
:MEASure:DUAL:VPP? [<source1>][,<source2>]
The :MEASure:DUAL:VPP? query measures the maximum and minimum vertical
value for the selected source, then calculates the vertical peak-to-peak value and
returns that value. The peak-to-peak value (Vpp) is calculated with the following
formula:
Vpp = Vmax - Vmin
Vmax and Vmin are the vertical maximum and minimum values present on the
selected source.
Return Format
<value><NL>
<value> ::= vertical peak to peak value in NR3 format
See Also
•
•
•
•
•
•
"Introduction to :MEASure Commands"
•
•
•