:MEASure Commands
21
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
473
:MEASure:DUAL:CHARge
Overview
This measurement is available with the N2820A high sensitivity current probe
when both the Primary and Secondary probe cables are used. This measurement
joins the Zoom In waveform data below the probe's clamp level with Zoom Out
waveform data above the probe's clamp level to create the waveform on which the
measurement is made.
Command Syntax
:MEASure:DUAL:CHARge [<interval>][,<source1>][,<source2>]
<interval> ::= {CYCLe | DISPlay}
<source1>,<source2> ::= CHANnel<n> with N2820A probe connected
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:DUAL:CHARge command installs a charge measurement on screen.
Charge measurements show the area between the waveform and the ground level.
The <interval> option lets you specify the measurement interval: either an integral
number of cycles, or the full screen. If <interval> is not specified, DISPlay is
implied.
If the optional source parameter(s) are specified, the currently specified source(s)
are modified.
Query Syntax
:MEASure:DUAL:CHARge? [<interval>][,<source1>][,<source2>]
The :MEASure:DUAL:CHARge? query measures and returns the charge
measurement value.
Return Format
<value><NL>
<value> ::= the charge value in Amp-hours in NR3 format
See Also
•
•
•
•
•
•
"Introduction to :MEASure Commands"
•