349
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes
SCPI Programmer's Guide
15 :EXTernal Trigger Commands
Control the input characteristics of the external trigger input. See
Introduction to
:EXTernal Trigger
Commands
The EXTernal trigger subsystem commands control the input characteristics of the
external trigger input. The probe factor, impedance, input range, input protection
state, units, and bandwidth limit settings may all be queried. Depending on the
instrument type, some settings may be changeable.
Reporting the Setup
Use :EXTernal? to query setup information for the EXTernal subsystem.
Return Format
The following is a sample response from the :EXTernal query. In this case, the
query was issued following a *RST command.
:EXT:BWL 0;RANG +8.0E+00;UNIT VOLT;PROB +1.000E+00
Table 90
:EXTernal Trigger Commands Summary
Command
Query
Options and Query Returns
:EXTernal:BWLimit
<bwlimit> (see
:EXTernal:BWLimit?
<bwlimit> ::= {0 | OFF}
:EXTernal:PROBe
<attenuation> (see
:EXTernal:PROBe? (see
<attenuation> ::= probe
attenuation ratio in NR3 format
:EXTernal:RANGe
<range>[<suffix>]
(see
:EXTernal:RANGe? (see
<range> ::= vertical full-scale
range value in NR3 format
<suffix> ::= {V | mV}
:EXTernal:UNITs
<units> (see
:EXTernal:UNITs? (see
<units> ::= {VOLT | AMPere}