296
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
10
:CHANnel<n> Commands
:CHANnel<n>:PROBe:RSENse
Command Syntax
:CHANnel<n>:PROBe:RSENse <value>
<value> ::= Ohms in NR3 format
<n> ::= 1 to (# analog channels) in NR1 format
When the N2820A high-sensitivity current probe is used with the N2825A
user-defined R-sense head, the :CHANnel<n>:PROBe:RSENse command specifies
the value of the R-sense resistor that is being probed in the device under test
(DUT). Suppoeted R-sense resistor values are from 1 m
Ω
to 1M
Ω
.
Query Syntax
:CHANnel<n>:PROBe:RSENse?
The :CHANnel<n>:PROBe:RSENse? query returns the R-sense resistor value
setting.
When the N2820A high-sensitivity current probe is not attached, the query returns
"INF".
When the N2820A high-sensitivity current probe is attached and a special head is
attached to the probe, the query returns the value of the special head. For
example, if the head is a "20 m
Ω
" head, the query returns 0.02.
Return Format
{<value> | INF}<NL>
<value> ::= Ohms in NR3 format
See Also
•
•
•
•
•
•
•