1158
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
31
:TRIGger Commands
:TRIGger:TV:UDTV:HTIMe
Command Syntax
:TRIGger:TV:UDTV:HTIMe <time>
<time> ::= seconds in NR3 format
When the Generic video trigger's horizontal sync control is enabled, the
:TRIGger:TV:UDTV:HTIMe command sets the minimum time the horizontal sync
pulse must be present to be considered valid.
Query Syntax
:TRIGger:TV:UDTV:HTIMe?
The :TRIGger:TV:UDTV:HTIMe query returns the horizontal sync time setting.
Return Format
<time><NL>
<time> ::= seconds in NR3 format
See Also
•
•