Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29
:SBUS<n> Commands
:SBUS<n>:LIN:TRIGger:PATTern:DATA
Command Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA <string>
<string> ::= "n" where n ::= 32-bit integer in unsigned decimal when
<base> = DECimal
<string> ::= "nn...n" where n ::= {0 | 1 | X | $} when
<base> = BINary
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $} when
<base> = HEX
The :SBUS<n>:LIN:TRIGger:PATTern:DATA command specifies the LIN trigger data
pattern searched for in each LIN data field.
Set a <string> bit to "0" or "1" to set the corresponding bit in the data pattern to
low or high, respectively.
Set a <string> bit to "X" to ignore (mask off) that bit in the data pattern.
Use the "$" character to indicate that the value of the corresponding bit will not be
changed (the existing bit value is used).
When <base> = DECimal, the "X" and "$" characters cannot be entered. When
queried, the "$" character is returned when any bits in the pattern have the value
of "X" and <base> = DECimal. When any bits in a given nibble have the value of "X"
and <base> = HEX, the "$" character is returned for the corresponding nibble.
Query Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA?
The :SBUS<n>:LIN:TRIGger:PATTern:DATA? query returns the currently specified
LIN trigger data pattern.
Return Format
<string><NL>
NOTE
<base> is specified with the :SBUS<n>:LIN:TRIGger:PATTern:FORMat command. The default
<base> is BINary.
NOTE
The length of the trigger data value is determined by the
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth command.
NOTE
If more bits are sent for <string> than the specified trigger pattern data length, the most
significant bits will be truncated. If the data length size is changed after the <string> is
programmed, the added or deleted bits will be added to or deleted from the least significant
bits.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...