Vertical Controls
3
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
69
For some active probes (such as InfiniiMax probes), the oscilloscope can
accurately calibrate its analog channels for the probe. When you connect a
probe that can be calibrated, the
Calibrate Probe
softkey appears (and the probe
attenuation softkey may change). See
See Also
•
"To specify the channel units"
•
"To specify the probe attenuation"
•
To specify the channel units
1
Press the probe's associated channel key.
2
In the Channel Menu, press
Probe
.
3
In the Channel Probe Menu, press
Units
; then, select:
•
Volts
— for a voltage probe.
•
Amps
— for a current probe.
Channel sensitivity, trigger level, measurement results, and math functions will
reflect the measurement units you have selected.
To specify the probe attenuation
This is set automatically if the oscilloscope can identify the connected probe. See
Analog channel inputs (see
The probe attenuation factor must be set properly for accurate measurement
results.
If you connect a probe that is not automatically identified by the oscilloscope, you
can manually set the attenuation factor as follows:
1
Press the channel key.
2
Press the
Probe
softkey until you have selected how you want to specify the
attenuation factor, choosing either
Ratio
or
Decibels
.
3
Turn the Entry knob
to set the attenuation factor for the connected probe.
When measuring voltage values, the attenuation factor can be set from 0.001:1 to
10000:1 in a 1-2-5 sequence.
When measuring current values with a current probe, the attenuation factor can
be set from 1000 V/A to 0.0001 V/A.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...