Acquisition Control
Using Peak Detect Mode to Find a Glitch
1
Connect a signal to the oscilloscope and obtain a stable display.
2
To find the glitch, press the
[Acquire]
key; then, press the
Acq Mode
softkey until
Peak Detect
is selected.
3
Press the
[Display]
key then press the
∞
Persistence
(infinite persistence) softkey.
Infinite persistence updates the display with new acquisitions but does not
erase previous acquisitions. New sample points are shown at normal intensity
while previous acquisitions are displayed at reduced intensity. Waveform
persistence is not kept beyond the display area boundary.
Press the
Clear Display
softkey to erase previously acquired points. The display
will accumulate points until
∞
Persistence
is turned off.
4
Characterize the glitch with Zoom mode:
a
Press the
zoom key (or press the
[Horiz]
key and then the
Zoom
softkey).
b
To obtain a better resolution of the glitch, expand the time base.
Figure 37
Sine With Glitch, Peak Detect Mode
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...