Troubleshooting
4
Infiniium V-Series Oscilloscopes Service Guide
75
Display Troubleshooting
Display troubleshooting flowchart
Connect external
monitor to VGA
port. Turn on power.
Does display
appear on ext.
monitor?
Is front panel
display
scrambled?
Is front panel
display black?
Check LED backlight cable,
front panel power cable, and
backlight enable connector
Still black?
Replace front panel
board
Still black?
Replace LCD
Still black?
Done
No
Yes
No
Yes
Yes
Replace
motherboard
Replace
motherboard
Yes
No
Yes
Check display cable
connection to
display and
motherboard
Yes
Still scrambled?
Yes
No
No
Replace
motherboard
NOTE
When you connect an external monitor the first time, it may display the start-up BIOS
information and then go blank when Windows starts up. If so, use the Windows settings to
enable the external monitor.