Replacing Assemblies
Removing and replacing the backlight inverter board
1
Disconnect the power cable and remove the cover and top plate.
2
Disconnect the cables from the backlight inverter board.
3
Remove two Torx T20 screws that connect the board to the chassis.
4
Lift the backlight inverter board out through the top of the chassis.
5
To replace the backlight inverter board, reverse this procedure.
WARNING
SHOCK HAZARD! The backlight inverter assembly, which is mounted at the front
corner of the oscilloscope near the flat-panel display, operates at high voltages
from 300-1 kVACrms. DO NOT handle this assembly while it is in operation.