Keysight Technologies E8267D Installation Notes Download Page 4

4

Installation Kit Parts List

Tools Required

TORX  T-10  driver

TORX  T-15  driver

TORX  T-20  driver

9/16” nut driver

5 mm nut driver

9/16” socket

5 mm socket

Torque Wrench Ratchet (9-21 in-lb)

Small Phillips screw driver

Needle-nose pliers

Standard screw driver

Safety Considerations

WARNING

Before you disassemble the signal generator, turn the power switch off and unplug the 
power cord. Failure to unplug the signal generator can result in personal injury.

CAUTION

Electrostatic discharge (ESD) can damage or destroy electronic components. All work on 
electronic assemblies should be performed at a static-safe workstation.

Item

Quantity

Description

Part Number

1

1

Installation Note

E8251-90388

2

1

W64 Ribbon Cable Assembly

8121-1328

3

1

W90 Digital Bus Cable

8121-0690

4

1

Hex Nut 15/21-32 (AUX I/O)

2950-0035

5

1

Washer .4721D (BNC)

2190-0102

6

1

A14 Digital Bus Baseband Generator 8/64 Msample

E4400-60765

a

a. Requires firmware version C.06.00 or greater. Visit: 

http://www.agilent.com/find/upgradeassistant

7

1

W65 Cable A33/A04 400G

8121-0643

8

1

Entitlement Certificate

5964-5143

9

1

Entitlement Certificate Envelope

5010-7743

Summary of Contents for E8267D

Page 1: ...sight Technologies For more in formation go to www keysight com Keysight E8267D PSG Signal Generators Upgrade Kit for Option 601 and 602 Digital Bus Baseband Generator with 8 and 64 Msample Memory res...

Page 2: ...this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for...

Page 3: ...2 Initialize the signal generator memory 3 Remove the outer and inner signal generator covers 4 Install the A14 Digital Bus Baseband Generator board 5 Verify proper installation of Option 601 602 6 R...

Page 4: ...Electrostatic discharge ESD can damage or destroy electronic components All work on electronic assemblies should be performed at a static safe workstation Item Quantity Description Part Number 1 1 In...

Page 5: ...2 Run the signal generator self test by pressing Utility Instrument Info Help Mode Self Test Run Complete Self Test Upon completion a summary of the self test will be displayed Use the E8257D 67D PSG...

Page 6: ...nter screws 3 on the four rear panel feet 4 4 Remove the four bottom feet 5 from the cover by lifting the tab and sliding the foot toward the tab 5 Place the signal generator on its side 6 Tilt the si...

Page 7: ...7 Figure 1 Outer Cover Removal...

Page 8: ...8 Figure 2 Inner Top Cover Removal...

Page 9: ...Figure 3 and Figure 4 for this procedure 1 Disconnect the power cord 2 Disconnect the W66 ribbon cable from A18P2 3 Using the standard screwdriver unlock the A41 door latch screw by rotating counter c...

Page 10: ...4 Connect the W64 ribbon cable to A17P2 5 Connect the W90 cable end with foam attached to it to A14J21 6 Connect the W90 cable to A17P1 7 Connect the W90 cable to A14J21 8 Connect the W64 ribbon cable...

Page 11: ...11 Figure 5 Baseband Cover Removal 1 1 1 1...

Page 12: ...12 Figure 6 A14 Internal Baseband Generator Board Figure 7 A17 Baseband Generator Interface Board A14 W65 A14J801 W64 A14J13 W90 A14J21 W64 A17P2 W90 A17P1...

Page 13: ...N Make sure the angled side of the cable connector on W90 is facing up Failure to do so could result in damage to the A17 Rear Panel Interface Board connector Figure 8 Digital Bus Cable Installation W...

Page 14: ...15 and T 20 screws to 21 in lbs 4 Attach the Digital Bus label over DIG I Q I O on the rear panel Run the I Q Calibration 1 Turn the instrument on and let it warm up for 30 minutes 2 Press I Q I Q Cal...