Setup for most tests
This setup is used for most tests and it requires the DMM, electronic load, and power supply being
verified. Some wire is also required for connection between instruments. A LAN or USB cable is needed
for readback data. The DMM measures the power supply output, and the electronic load draws current
from the power supply. The accuracy of the current monitoring resistor must be 0.01% or better, which
should include any self-heating effects.