*OPC
*OPC?
Sets the “Operation Complete” bit (bit 0) of the Standard Event register after the command is executed.
The query returns 1 to the output buffer after the command is executed.
*OPT?
Returns the option number of the instrument.
*PSC 0|1
*PSC?
(
Power- on status clear
) Enables (1) or disables (0) the clearing of the Status Byte and the Standard
Event register enable masks when power is turned on.
*RST
Resets the instrument to its default state, but does not clear any of the status registers or the error
queue. It also does not affect any interface error conditions.
Parameter
Default Value
OCP delay
50 ms
OVP level
Maximum (varies by model)
Voltage
0
Current
Maximum (varies by model)
OCP state
Off
OVP state
Off
Output state Off
Relay sense Internal
*RCL <
state
>
*SAV <
state
>
Saves (*SAV) the current instrument state or recalls (*RCL) a previously saved state file. All state storage
locations (0 through 9) are nonvolatile.
– The parameters stored in a state file are the same parameters affected by *RST.
– States are not affected by the *RST command.
– Saving a state overwrites the previous state (if any) stored in that location.
– When shipped from the factory, storage locations 0 through 9 are empty.