04 | Keysight | Using an External Trigger to Measure Current with B2980A Series – Demo Guide
Setup
Connection
1. Connect the
Common terminal
to the
Earth terminal
of the B2985A/87A with a
banana to
screw-lug cable
.
2. Connect the
Voltage Source Low terminal
to the
Common terminal
of the B2985A/87A with a
black banana test lead
.
3. Connect the
Ammeter Input terminal
of the B2985A/87A to the
J1 terminal
of the N1422A with
a
triaxial cable
.
4. Connect the
Voltage Source High terminal
of the B2985A/87A to the
J2 terminal
of the N1422A
with a
red banana test lead
.
5. Connect the
trigger source
to the
Trigger In port
of the B2985A/87A with a
BNC cable
.
N1422A
B2985A/87A Terminals