11 | Keysight | Using an External Trigger to Measure Current with B2980A Series – Demo Guide

The 

current measurement range operation

 is set to 

AUTO

 in the default settings. When 

the operation is set to AUTO, an appropriate current measurement range is selected 
automatically by the instrument depending on the current measurement result. If the 
mode is set to FIXED, the current measurement range is fixed to the one specified by the 
user during the measurement.

1) Press            to hide 

Trigger Parameters

.

2) Press               to set the 

current measurement range operation

 to 

FIXED

.

3) Press the measurement navigation key                     to select a 

larger current 

measurement range

.

4) Press the measurement navigation key                     to select a 

smaller current

measurement range

.

8. (Optional) If you’d like to change the measurement range operation

Summary of Contents for B2980A Series

Page 1: ...Keysight Technologies Using an External Trigger to Measure Current with B2980A Series B2980A Femto Picoammeter and Electrometer High Resistance Meter Demo Guide...

Page 2: ...ctrometers the B2980A series features a 4 3 color LCD graphical user interface GUI enabling various viewing capabilities such as numeric graph histogram and roll trend chart which allows you to easily...

Page 3: ...l need to configure the B2985A 87A ammeter to accept an external trigger and make a current measurement Figure 3 shows the external trigger signal requirements to initiate a current measurement from t...

Page 4: ...ce Low terminal to the Common terminal of the B2985A 87A with a black banana test lead 3 Connect the Ammeter Input terminal of the B2985A 87A to the J1 terminal of the N1422A with a triaxial cable 4 C...

Page 5: ...ernal trigger signal 6 Optional Abort the ammeter to stop waiting for an external trigger signal 7 Optional If you d like to change aperture mode 8 Optional If you d like to change measurement range o...

Page 6: ...ss to edit it Then enter 1 s to set the Measure trigger period to 1 s which should be longer than the sum of the measurement delay time and the aperture time 5 Rotate to select the Measure trigger sou...

Page 7: ...ave your settings 10 Rotate to select the Measure trigger count and press to edit it Then enter the Number of trigger signals which will be sent to the instrument as the Measure trigger count 11 Or pr...

Page 8: ...current across it 1 Rotate to select the Voltage source and press to edit it 2 Then rotate until the Voltage Source value reaches 1 V and press to fix it 3 Press Voltage Source to enable Voltage Sourc...

Page 9: ...end of the test 1 Press and then press to make the ammeter stop waiting for an external trigger signal 7 Optional If you d like to change the aperture mode The automatic aperture mode is selected in t...

Page 10: ...ment navigation key to select longer aperature type 3 For example if you d like to select the manual aperture mode press 4 If you d like to set the mode to the automatic aperture mode again press 5 If...

Page 11: ...g on the current measurement result If the mode is set to FIXED the current measurement range is fixed to the one specified by the user during the measurement 1 Press to hide Trigger Parameters 2 Pres...

Page 12: ...in the B2980A Series to wait for an external trigger signal The series of commands shown in group D E and F are optional to abort the ammeter to stop waiting for an external trigger signal to change t...

Page 13: ...easurement features that can maximize your measurement confidence The 0 01 fA 10 17 A minimum current resolution is available for both Femto Picoammeters and Electrometers High Resistance Meters and s...