3-12
Keysight B2900 User’s Guide, Edition 5
Installation
Maintenance
•
For preventive maintenance
To perform
self-test
The following procedure shows how to perform the self-test.
1. If it is in the remote state, press the Cancel / Local key.
2. Press the On/Off switch and confirm that the switch turns off.
3. Disconnect test leads and cables from the channel terminals.
4. Press the following function keys.
More > System > Cal/Test > Self-Test
Confirmation dialog box is opened.
5. Press the OK key. This starts the self-test.
Self-calibration
Keysight B2900 provides the self-calibration function to maintain the measurement
performance. If the environmental temperature changes
3
C or more, perform the
self-calibration. This is effective for the accurate measurements by minimizing the
effect of thermal drift. The self-calibration must be performed after warming-up of
60 minutes. Before performing the self-calibration, turn the channel output off and
disconnect test leads and cables from the terminals.
To perform
self-calibration
The following procedure shows how to perform the self-calibration.
1. If it is in the remote state, press the Cancel / Local key.
2. Press the On/Off switch and confirm that the switch turns off.
3. Disconnect test leads and cables from the channel terminals.
4. Press the following function keys.
More > System > Cal/Test > Self-Cal
Confirmation dialog box is opened.
5. Press the OK key. This starts the self-calibration.
Summary of Contents for B2900 Series
Page 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Page 3: ......
Page 25: ...1 Getting Started...
Page 41: ...2 Introduction...
Page 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Page 65: ...3 Installation...
Page 101: ...4 Front Panel Reference...
Page 155: ...5 Front Panel Operations...
Page 189: ...6 Function Details...
Page 233: ......