6-20
Keysight B1542A User’s Guide, Edition 10
PLSDIV TIS Commands
Command Reference
PlsdivSetMeasTime
This command changes the timing for extracting the pulsed IV measurement data
(gate voltage, drain voltage, and drain current) from the oscilloscope measurement
result (pulse waveform). This change is effective after the PlsdivForceVg command
is executed and until the next PlsdivSetMeasTime command or the PlsdivUninit
command is executed. This command execution is optional.
Remarks
Changing this value may cause measurement errors.
Filename
PlsdivSetMeasTime.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Write String
This command needs the following input parameter.
meas_time
Data extraction timing. Numeric value,
-
1 to 1. Default value is
0.5. The data extraction timing is specified by the rate for the
pulse width as shown in
.
Figure 6-5
Data Extraction Timing
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
offset = 0
offset = 0
Gate pulse
Drain pulse
Data extraction timing
meas_time
x Pulse width
V/2
V/2
Pulse width
V: Pulse amplitude, > 5.5 mV, < 4.5 V
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......