Keysight B1505A Configuration and Connection Guide
5-25
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect High Voltage Bias Tee
To Connect High Voltage Bias Tee
Keysight N1260A high voltage bias-tee is used to perform the high voltage C-V
measurement. The C-V measurement of up to 3000 Vdc can be realized by using the
bias-tee, MFCMU, and HVSMU. The N1260A should be installed near your DUT
interface as shown in
The N1260A is also used for the high voltage C-V measurement using the N1265A test
fixture. It must be connected between the MFCMU and the N1265A. Then use two
N1254A-518 SHV cables to connect the N1260A output to the N1265A Bias Tee input.
The N1260A input connection is same as
Table 5-18
To Connect N1260A Input
NOTE
Turn the B1505A off before connecting the CMU cable.
Connect HVSMU to obtain DC bias input. Because the bias-tee cannot pass the MFCMU
internal DC bias.
B1505A terminals
Required cables N1260A
terminals
HVSMU
Force
16493T HVSMU cable, 3 m or 1.5 m
From HVSMU
MFCMU
Hcur
N1300A CMU cable, 3 m or 1.5 m
It is not needed to connect the ground wire extended from the
CMU cable.
From
MFCMU
Hcur
Hpot
Hpot
Lpot
Lpot
Lcur
Lcur
From HVSMU
Hcur
Hpot
Lpot
Lcur
From MFCMU
㼼
3 k V M a x
䞉䞉䞉䞉
㼼
3 k V M a x
䞉䞉䞉䞉䞉䞉䞉䞉
!!
Summary of Contents for B1505A
Page 3: ......
Page 9: ...1 Configuration Guide ...
Page 83: ...2 N1259A Connection Guide ...
Page 107: ...3 N1265A Connection Guide ...
Page 133: ...4 N1272A and N1273A Connection Guide ...
Page 149: ...5 Connection Guide for Wafer Prober and Your Own Test Fixture ...
Page 203: ...6 Accessory Dimensions ...
Page 219: ...7 Connection and Ordering Examples ...
Page 297: ......