2-20 Keysight 85309B User’s and Service Guide
Operating Information
Performing the Operator’s Check
2-
a.
Place the 85309B front panel Band Selection switch in the HIGH
position, and compare the output power traces with the limits
across the entire frequency range in
b.
Place the 85309B front panel Band Selection switch in the LOW
position, and compare the output power traces with the limits
across the entire frequency range in
34.
Set the PNA Port 1 source power to 6 dBm.
35.
For Option 002 and 402 Models Only: Checking the 85309B Test 3
LO/IF (J13) Channel
36.
Connect a test cable between the PNA Port 1 and the 85309B rear panel
port LO Input (J1).
37.
Connect a test cable between the PNA Port 2 and the 85309B rear panel
port Test 3 LO/IF (J13).
38.
Set the PNA Port 1 source power to 0 dBm.
39.For Option 002 models only:
Compare the output power traces from the
Test 3 LO/IF (J13) port with the limits across the entire frequency range in
40.For Option 402 models only:
a.
Place the 85309B front panel Band Selection switch in the HIGH
position, and compare the output power traces with the limits
across the entire frequency range in
b.
Place the 85309B front panel Band Selection switch in the LOW
position, and compare the output power traces with the limits
across the entire frequency range in
41.
Set the PNA Port 1 source power to 6 dBm.
42.
Table 2-3
85309B Standard or Options 001 or 002 Output Power
Limits
LO Output Power (+6 dBm Input)
Frequency Range
Power Level
0.3 – 0.5 GHz
> +21.6 dBm
0.5 – 3 GHz
> +24.1 dBm
3 – 6.2 GHz
> +26.3 dBm
6.2 – 18 GHz
> +23.5 dBm
LO Output Power (0 dBm Input)
Summary of Contents for 85309B LO/IF
Page 6: ...vi Keysight 85309B User s and Service Guide...
Page 18: ...1 10 Keysight 85309B User s and Service Guide Getting Started Getting Help with Your 85309B 1...