3-16
Keysight Technologies 85059A 1.0 mm Calibration and Verification Kit
Making Connections
Using the Torque Wrench
Figure 3-10 Where to Hold the Torque Wrench
3. Hold the torque wrench lightly at the end of the handle
only
(beyond the
Figure 3-11 Using the Torque Wrench
4. Apply force perpendicular to the wrench handle. This applies torque to the
connection
through
the wrench.
Do
not
hold the wrench so tightly that you push the handle straight down along
its length rather than pivoting it. Doing so may apply an unlimited amount of
torque.
5. Tighten the connection just to the torque wrench “break” point (see
). Do
not
tighten the connection further.