Keysight 85033E User’s and Service Guide 3-
9
Use, Maintenance, and Care of the Devices
Gaging Connectors
a.
While holding the gage by the barrel, and without turning the gage
or the device, screw the connecting nut of the male device being
measured onto the male gage, just until you meet resistance.
Connect the nut finger-tight. Do not overtighten.
b.
Use the torque wrench recommended for use with the kit to tighten
the connecting nut to 90 N-cm (8 in-lb). Refer to
for more information.
c.
Gently tap the barrel of the gage with your finger to settle the gage
reading.
d.
Read the gage indicator dial. Read
only
the black
signs;
not
the
red
signs.
For maximum accuracy, measure the connector a minimum of three
times and take an average of the readings. After each measurement,
rotate the gage a quarter-turn to reduce measurement variations
that result from the gage or the connector face not being exactly
perpendicular to the center axis.
e.
Compare the average reading with the observed pin depth limits in