214
8163A/B, 8164A/B & 8166A/B Mainframes Programming Guide
4
Measurement Operations & Settings
command:
:TRIGger[
n
][:CHANnel[
m
]]:OUTPut:REARm
syntax:
:TRIGger[
n
][:CHANnel[
m
]]:OUTPut:REARm<wsp>OFF|ON|0|1
description:
Sets the arming response of a channel to an outgoing trigger.
See
on page -209, for information on how this command affects the generation of output triggers using power
measurements.
parameters:
A
boolean
value:
OFF
or
0
: trigger rearming disabled
ON
or
1
: trigger rearming enabled (default)
If you return to Local control, all modules return to the default setting.
response:
none
example:
trig1:outp:rearm 1
affects:
All power meters, and return loss modules.
dual sensors:
Can only be sent to master channel, slave channel is also affected.
command:
:TRIGger[
n
][:CHANnel[
m
]]:OUTPut:REARm?
syntax:
:TRIGger[
n
][:CHANnel[
m
]]:OUTPut:REARm?
description:
Returns the arming response of a channel to an outgoing trigger.
parameters:
none
response:
A
boolean
value:
0
: trigger rearming disabled (default)
1
: trigger rearming enabled.
example:
trig1:outp:rearm?
->
0<END>
affects:
All power meters, and return loss modules.
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
command:
:TRIGger:CONF
syntax:
:TRIGger:CONF<wsp><triggering mode>
description:
Sets the hardware trigger configuration with regard to Output and Input Trigger Connectors.