X, P, and K 11644A
4-5
Use, Maintenance, and Care of the Devices
Connections
Connections
Good connections require a skilled operator. Slight errors in operator technique can have a significant
effect on measurements and measurement uncertainties
. The most common cause of measurement error
is poor connections.
The following procedures illustrate how to make good connections.
Waveguide Devices
IMPORTANT
Unlike threaded devices, the WR-90, WR-62, WR-42 waveguide mating planes are flanges
(often precision) that you must carefully screw together. Always connect waveguide in the
same flange orientation. For example, use the label as a reference and always connect a
device with the label facing the same direction.
Precision Flanges
A precision flange has four corner screw holes
and
two precision alignment holes, as shown in
.
A non-precision flange has only four corner screw holes
.
Figure 4-2 Precision Alignment Holes
Summary of Contents for 11644A K
Page 2: ......
Page 3: ...X P and K 11644A i...
Page 7: ...1 General Information...
Page 12: ...1 6 X P and K 11644A General Information Preventive Maintenance...
Page 13: ...2 Specifications...
Page 19: ...3 User Information...
Page 26: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition...
Page 27: ...4 Use Maintenance and Care of the Devices...
Page 36: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage...
Page 37: ...5 Performance Verification...
Page 52: ...5 16 X P and K 11644A Performance Verification System Operation Checks...
Page 53: ...6 Troubleshooting...
Page 57: ...7 Replaceable Parts...
Page 58: ...7 2 X P and K 11644A Replaceable Parts...
Page 67: ...A Standard Definitions...