11
(1) Probe selection and attachment
Check that the power supply is off and then attach either
the LEP-J or LHP-J probe according to the substrate to be
measured (see p. 10).
(2) Power ON
Press the Power key.
(3) Calibration
Check whether the measurement target has already been
calibrated.
If not, then perform calibration and register the
calibration curve (application) (see p.24).
(4) Measurement
The probes are of the "one-point contact constant-pressure
type," whereby a constant load is applied to the tip of the probe.
As shown in the figure on the right, hold the part near the
measuring part and lower the probe rapidly so that the probe
is at a right angle in relation to the measurement surface. To
prepare for the next measurement, lift off the probe tip once at
least 10 mm from the measurement surface.
When measuring a cylindrical object or continuously
measuring a flat surface, the use of a probe adapter allows
stable measurement.
Probe usage method
Probe adapter usage
method
Attach the probe
adapter to the
tip of the probe
and secure it by
tightening the
screw.
Rapidly lower the probe onto
the measurement surface.
Display Example (Electromagnetic Measurement Mode)
Measurement enabled state
Measurement result
Measurement
6. Operation Method