20
* At least two samples and up to eight samples are needed. Ar-
range the results in ascending order of the D mode values as
shown in the [Table 1]. Each point is referred to as a zigzag
point, and connect zigzag points with a straight line. For points
less than the minimum point and more than the maximum point
of the calibrations, a prolongation of the adjacent line is applied.
In the case of the graph at right, the prolongation of
①
−
②
is
applied to the part below
①
, and the prolongation of
③
−
④
is ap-
plied to the part above
④
.
* The user calibrations can also be created in the S mode as well
as the D mode.
* The use of data logger software, “HDL-01”, (option) allows users
to simplify user calibrations.
Eg.) When the D mode of 246 and the moisture content
of 12.4 are entered into user calibration number 1:
(1) Select “
Make User Cal
” on the extension function and press the
key.
(2) Use the
▲
or
▼
key to select “
User Cal. No 1
” and press the
key.
D mode Moisture
value
①
246
12.4
②
688
26.7
③
1015
44.2
④
1578
52.3
Table 1